WCDMA Signaling
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
491
User Manual 1173.9657.02 ─ 11
CONFigure:WCDMa:SIGN<i>:BER:SCONdition
<StopCondition>
Qualifies whether the measurement is stopped after a failed limit check or continued.
SLFail
means that the measurement is stopped and reaches the
RDY
state as soon
as one of the results exceeds the limits.
Parameters:
<StopCondition>
NONE | SLFail
NONE
: Continue measurement irrespective of the limit check
SLFail
: Stop measurement on limit failure
*RST:
NONE
Example:
See
Configuring the BER Measurement
Firmware/Software:
V1.0.15.0
Manual operation:
See
CONFigure:WCDMa:SIGN<i>:BER:TBLocks
<TransportBlocks>
Defines the number of transport blocks to be measured per measurement cycle (statis-
tics cycle).
Parameters:
<TransportBlocks>
Range:
1 to 50E+3
*RST:
100
Example:
See
Configuring the BER Measurement
Firmware/Software:
V1.0.15.0
Manual operation:
See
CONFigure:WCDMa:SIGN<i>:BER:PNResync
<Enable>
Activates or deactivates a correction (reordering) mechanism for transports blocks loo-
ped back in wrong order.
Parameters:
<Enable>
OFF | ON
ON
: Correction meachanism active, BER measurement result
based on corrected block sequence, number of corrected blocks
available as result "PN Discontinuity"
OFF
: Correction meachanism inactive, no "PN Discontinuity"
result
*RST:
ON
Example:
See
Configuring the BER Measurement
Firmware/Software:
V1.0.15.0
Manual operation:
See
Command Reference