WCDMA Signaling
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
242
User Manual 1173.9657.02 ─ 11
Results
For a detailed description of the results see
chapter 2.2.18.2, "Measurement Results"
Remote command:
2.4.22.3
Measurement Control Settings
The "Measurement Control" parameters configure the scope of the measurement.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "System Over-
view"
Fig. 2-119: Measurement control settings
Repetition
Defines how often the measurement is repeated if it is not stopped explicitly or by a
failed limit check.
●
Continuous:
The measurement is continued until it is explicitly terminated; the
results are periodically updated.
●
Single-Shot:
The measurement is stopped after one statistics cycle.
Single-shot is preferable if only a single measurement result is required under fixed
conditions, which is typical for remote-controlled measurements. Continuous mode is
suitable for monitoring the evolution of the measurement results in time and observe
how they depend on the measurement configuration, which is typically done in manual
control. The reset/preset values therefore differ from each other.
Remote command:
Stop Condition
Specifies the conditions for an early termination of the measurement:
●
None:
The measurement is performed according to its "Transport Blocks", irre-
spective of the limit check results.
●
On Limit Failure
: The measurement is stopped as soon as one of the limits is
exceeded. If no limit failure occurs, it is performed according to its "Transport
GUI Reference