WCDMA Signaling
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
250
User Manual 1173.9657.02 ─ 11
Repetition
Defines how often the measurement is repeated if it is not stopped explicitly or by a
failed limit check.
●
Continuous:
The measurement is continued until it is explicitly terminated; the
results are periodically updated.
●
Single-Shot:
The measurement is stopped after one statistics cycle.
Single-shot is preferable if only a single measurement result is required under fixed
conditions, which is typical for remote-controlled measurements. Continuous mode is
suitable for monitoring the evolution of the measurement results in time and observe
how they depend on the measurement configuration, which is typically done in manual
control. The reset/preset values therefore differ from each other.
Remote command:
Measure Frames
Specifies the number of subframes to be measured per measurement cycle (statistics
cycle).
Remote command:
Limit
Defines an upper limit for the E-HICH reception "False Ratio" result.
Remote command:
2.4.25 RLC Throughput Measurement Configuration
The signaling RLC Throughput measurement is included in the WCDMA signaling
application. It must be activated in the "Measurement Controller" dialog via the entry
"RX Measurement...". The "RLC Throughput" tab of the RX measurement view and the
related configuration dialog are described in this section.
See also
chapter 2.2.21.1, "Performing RLC Throughput Measurements"
2.4.25.1
Measurement Control
The measurement is turned on or off using the ON | OFF or RESTART | STOP keys.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "System
Overview"
GUI Reference