R&S AFQ100A
Baseband Block
1401.3084.32 4.15
E-3
Repeat Count - Trigger
(Trigger mode Repeat Count only)
Sets a repeat rate for restarting the waveform after a trigger event
occurs. The waveform can be repeated up to 100 times, provided that
Enable Retrigger
is set (see
, on page
Remote-control command:
SOUR:TRIG:RCO 10
Enable Retrigger -
Trigger
Selects whether a following trigger event resets the current signal
output. The minimum time interval between two retrigger events
amounts to 10
W
s plus 25 memory clock symbol periods.
Remote-control command:
SOUR:TRIG:REN ON
Note:
For multi segment waveforms Enable Retrigger is always on.
Source - Trigger
Selects the trigger source.
Switching between the segments of a multi segment waveform is
triggered by a segment trigger. The segment trigger is configured
separately in the
Segment Trigger Configuration
section of the
dialog.
Note:
If the instrument is connected with further R&S AFQ I/Q Modulation
Generators all instruments which are set to
Slave
and the
Last
Slave
can be triggered
External
only (Synchronization mode see
Sync Mode - Reference Clock
, on page 4.34).
Manual
(Sync mode Master and Normal only)
The trigger event is executed by the
Execute
button or by pressing the
TRIGGER
key on the
front panel.
Remote-control command:
SOUR:TRIG:SOUR MAN
External
The trigger event is executed with the aid of the
active edge of an external trigger signal. The
trigger signal is fed in via the TRIG connector on
the rear panel. The active slope is selected under
Slope
.
Remote-control command:
SOUR:TRIG:SOUR EXT
Bus
(Sync mode Master and Normal only)
The trigger event is executed by remote control