![R&S 1444.1190K02 Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881069.webp)
Remote control commands
R&S
®
NRPxxP
69
User Manual 1179.5760.02 ─ 02
Parameters:
<lreference>
Specified in percent of the pulse amplitude.
Range:
0.0 to 100.0
*RST:
10.0
Default unit: PCT
8.6.5.4
High-resolution measurements
SENSe<Sensor>:TRACe:ESAMpling:AUTO
....................................................................... 69
SENSe<Sensor>:TRACe:MEASurement:TRANsition:ESAMpling:AUTO
................................ 69
SENSe<Sensor>:TRACe:ESAMpling:AUTO
<value>
Available in trace mode.
Enables or disables the automatic equivalent sampling that allows for high-resolution
measurements.
The automatic equivalent sampling requires a periodic measurement signal. The power
sensor superposes several successive measurements to improve the resolution. The
measurement speed decreases, because several successive measurements are nec-
essary.
Parameters:
<value>
ON | OFF
*RST:
ON
SENSe<Sensor>:TRACe:MEASurement:TRANsition:ESAMpling:AUTO
<val>
Available in pulse analysis measurements.
Enables or disables the automatic equivalent sampling that allows for high-resolution
measurements.
The automatic equivalent sampling requires a periodic measurement signal. The power
sensor superposes several successive measurements to improve the resolution. The
measurement speed decreases, because several successive measurements are nec-
essary.
Parameters:
<val>
OFF | ON
*RST:
ON
8.6.5.5
Transferring measured pulse parameters
SENSe<Sensor>:TRACe:MEASurement:AUTO
<mode>
Available in continuous average mode.
Enables or disables the automatic transfer of the measured pulse parameters after
each trace.
Configuring the measurement modes