![R&S 1444.1190K02 Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881067.webp)
Remote control commands
R&S
®
NRPxxP
67
User Manual 1179.5760.02 ─ 02
ON
The power sensor automatically determines the pulse parame-
ters for the currently measured trace.
OFF
Pulse analysis measurement disabled.
*RST:
OFF
8.6.5.2
Setting the duration
SENSe<Sensor>:TRACe:MEASurement:TIME
<time>
Available in pulse analysis measurements.
Sets the duration of the pulse analysis. The start time of the pulse analysis is set by
SENSe<Sensor>:TRACe:MEASurement:OFFSet:TIME
. The end time of the pulse
analysis is set by this command in relation to the end of the trace measurement.
Parameters:
<time>
A value of 0.0 means that the pulse analysis ends with the end
of the trace measurement.
A value larger than 0.0 shortens the duration of the pulse analy-
sis accordingly.
Range:
0.0 to 10.0
*RST:
0.0
Default unit: s
8.6.5.3
Defining reference levels for pulse analysis measurements
Define the reference levels for the pulse timing. All values are specified in percent of
the pulse amplitude.
Remote commands:
SENSe<Sensor>:TRACe:MEASurement:ALGorithm
............................................................67
SENSe<Sensor>:TRACe:MEASurement:DEFine:DURation:REFerence
................................ 68
SENSe<Sensor>:TRACe:MEASurement:DEFine:TRANsition:HREFerence
........................... 68
SENSe<Sensor>:TRACe:MEASurement:DEFine:TRANsition:LREFerence
............................ 68
SENSe<Sensor>:TRACe:MEASurement:ALGorithm
<algorithm>
Sets the analysis algorithm for detecting the pulse top and the pulse base power of a
pulsed signal. From these two power levels, the power sensor derives the reference
levels, using the probability density of the recorded trace. If the quality of the signal is
too low to identify the pulse top and the pulse base power, the power sensor uses the
maximum and minimum values of the trace instead.
Parameters:
<algorithm>
HISTogram | INTegration | PEAK
Configuring the measurement modes