Remote control commands
R&S
®
NRPxxP
57
User Manual 1179.5760.02 ─ 02
period to the size of the sampling window. Otherwise, the modulation can have a con-
siderable influence, even if the sampling window is much larger than the modulation
period.
8.6.1.4
Calculating the measurement time
Normally, the measurement time is calculated as follows:
MT = 2 * AC * APER + (2 * AC - 1) * 5
μ
s
With:
MT: overall measurement time
AC: average count
APER: aperture time
5
μ
s is the time for switching the chopper phase.
8.6.1.5
Configuring the result buffer
SENSe<Sensor>:[POWer:][AVG:]BUFFer:CLEar
.................................................................57
SENSe<Sensor>:[POWer:][AVG:]BUFFer:COUNt?
..............................................................57
SENSe<Sensor>:[POWer:][AVG:]BUFFer:SIZE
...................................................................57
SENSe<Sensor>:[POWer:][AVG:]BUFFer:STATe
.................................................................58
SENSe<Sensor>:[POWer:][AVG:]BUFFer:CLEar
Clears the contents of the result buffer.
Example:
BUFF:CLE
Usage:
Event
SENSe<Sensor>:[POWer:][AVG:]BUFFer:COUNt?
Queries the number of results that are currently stored in the result buffer.
Example:
BUFF:COUN?
Usage:
Query only
SENSe<Sensor>:[POWer:][AVG:]BUFFer:SIZE
<count>
Sets the size of the result buffer.
You can enable the buffer using
SENSe<Sensor>:[POWer:][AVG:]BUFFer:STATe
.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
Configuring the measurement modes