Measurement Basics
R&S
®
NRX
45
User Manual 1178.5566.02 ─ 07
The sensor assignment is deleted by a preset, reset or sanitization. If the R&S NRX
has no memory of a previous sensor assignment, the R&S NRX assigns the measure-
ments according to the port, to which the power sensors are connected. The number of
measurement panes is adapted automatically.
Example: Connecting sensors after a preset
1. Connect an R&S
NRQ6 to port
A.
The measurement results are displayed in measurement pane
1.
2. Connect an R&S
NRP33SN to port
C.
The display layout is extended to 3 measurement panes. The measurement results
of the R&S
NRP33SN are displayed in measurement pane
3. Measurement pane
2
is unused.
3. Disconnect the R&S
NRP33SN from port
C and connect it to port
B.
4. Press [Preset].
The display layout is reduced to 2 measurement panes. The power sensors are
assigned according to their port. The assignment of the R&S
NRQ6 is unchanged,
but the R&S
NRP33SN is assigned to measurement pane
2.
Further information:
●
Chapter 13.3, "Addressing Measurements and Sensors"
Suffix usage in remote control
5.3
Performing a Measurement
This measurement description is designed to give you a first impression. For further
information, see the description of the measurements, their results and their settings:
●
Chapter 6, "Configuration for All Measurement Types"
●
Chapter 7, "Measurement Types and Result Displays"
●
Chapter 8, "Sensor Configuration"
Setup
1. Connect one or more R&S power sensors to the R&S NRX. See
How many R&S power sensors you can connect depends on the options of your
R&S NRX. See
Chapter 5.1, "Parallel Measurements"
2. Connect each R&S power sensor to a DUT (signal source). See the user manual of
the R&S power sensor for information on topics that need your special attention.
Starting a measurement
1. Preset the R&S NRX and the connected R&S power sensors.