Remote Control Commands
R&S
®
NRX
384
User Manual 1178.5566.02 ─ 07
MEMory:STATe:DEFine
<register_name>[, <register_number>]
MEMory:STATe:MAP
<register_name>[, <register_number>]
Assigns a name to the setup stored in the memory location.
Take care to use recognizable names. The R&S NRX does not check whether a name
is unique. If you want to reset the factory default, use
.
The query returns the memory location that is assigned to the setup name.
Parameters:
<register_name>
Setup name; allowed are alphanumeric characters and special
characters.
<register_number>
Memory location of the setup
0
Factory-set setup, cannot be changed
1 to 19
Available memory locations
Example:
MEM:STAT:MAP "test",5
Example:
MEM:STAT:MAP? "test"
Manual operation:
See
See
MEMory:STATe:RESet
Resets the setup names to factory default, "Setup 1", "Setup 2" and so on.
Usage:
Event
MEMory:TABLe:DATA?
Queries the content of the selected table.
Use
The response consists of data pairs (frequency - offset):
<frequency 1>,<offset 1>,<frequency 2>,<offset 2>,<frequency 3>,<offset 3>, ...
Frequency in Hz, offset in dB.
Example:
MEM:TABL:DATA?
Query
1.02,0.00,2.03,
0.00,3.04,0.00
Response
Usage:
Query only
Managing Setups and Correction Tables
www.allice.de
Allice Messtechnik GmbH