Remote Control Commands
R&S
®
NRX
379
User Manual 1178.5566.02 ─ 07
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
CALibration<Sensor>:TEST:REFerence?
Requires a thermal power sensor, equipped with an additional test heater.
Queries the reference value of the power difference in W when measuring with the
external heater enabled and disabled. The reference value is determined during the
calibration process and stored in the calibration data set.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
DIAGnostic:INFO:OTIMe?
Queries the count of the built-in elapsed-time meter. The count is always output in
hours [h] and cannot be changed.
Usage:
Query only
TEST:DEVice:RESult?
Queries the test results of
Usage:
Query only
TEST:DEVice[:ALL]
[<argument>]
Performs tests for the keyboard, display and touch panel.
Parameters:
<argument>
The tests can be performed as single tests or as combined test.
"SubSystemGui:KeyboardTest"
Keyboard test
"SubSystemGui:DisplayTest"
Display test
"SubSystemGui:TouchTest"
Touch test
Example:
TEST:DEV "SubSystemGui:
KeyboardTest;DisplayTest;TouchTest";*OPC
Performs a combined test.
Running Selftests