Remote Control Commands
R&S
®
NRX
353
User Manual 1178.5566.02 ─ 07
R&S NRX parameter
Remote control command
Value
Upper power ratio limit
CALCulate<Measurement>:
LIMit<DirectionalChannel>:UPPer[:DATA]:
RATio[:VALue]
1.0
dB
Lower power ratio limit
CALCulate<Measurement>:
LIMit<DirectionalChannel>:LOWer[:DATA]:
RATio[:VALue]
1.0
dB
13.7.7
Frequency Selective Power Sensors
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:REFio:FREQuency
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:REFio:OUTPut[:STATe]
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:SOURce
............................. 354
[SENSe<Sensor>:]BANDwidth[:RESolution]:TYPE:AUTO[:STATe]
...................................... 354
[SENSe<Sensor>:]BANDwidth[:RESolution]:TYPE[:VALue]
................................................354
[SENSe<Sensor>:]BANDwidth[:RESolution][:VALue]
......................................................... 355
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:REFio:FREQuency
<freq>
Effective for:
●
R&S
frequency selective power sensors
●
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:SOURce
REFio
is set.
Sets the frequency of the reference clock signal that is supplied at the REF connector
of the power sensor.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<freq>
Range:
1.0e+7 to 1.2e+8
*RST:
1.0e+7
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:REFio:OUTPut[:
STATe]
<state>
Effective for R&S
frequency selective power sensors.
If the REF connector of the power sensor is used as an output, enables or disables the
output signal.
Suffix:
<Measurement>
.
1 to 8
Measurement
Configuring Sensors