Remote Control Commands
R&S
®
NRX
323
User Manual 1178.5566.02 ─ 07
FREQ2
Lower sampling rate
Recommended to avoid measurement errors caused due to ali-
asing effects. However, this setting extends the measurement
time.
*RST:
FREQ1
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>][:POWer]:BURSt:DTOLerance
<value>
Effective for burst average measurements.
Detects the falling edge of a burst. If the power keeps low for at least the set time, the
end of the burst is assumed. Modulation-specific power drops that are shorter than the
set value are ignored.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
0.0 to 0.3
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:APERture[:
VALue]
<value>
Effective for continuous average measurements.
Sets the width of the sampling window.
When measuring modulated signals, the measurement can show fluctuation due to the
modulation. If that is the case, adapt the size of the sampling window exactly to the
modulation period to get an optimally stable display. If the modulation period varies or
is not precisely known, you can also set
CHANnel<Channel>][:POWer][:AVG]:SMOothing[:STATe]
ON
.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Configuring Sensors