Remote Control Commands
R&S
®
NRX
284
User Manual 1178.5566.02 ─ 07
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit:DETect
<value>
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LIMit:DETect
<value>
Effective if the Out
1
/
Trig
Out connector is configured as monitoring output for the
power indication, for example using
[SENSe<Sensor>:]POWer[:POWer]:RANGe:
ON
.
Defines when a logic high level (> 2.7
V) is output at the Out
1
/
Trig
Out connector.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
INBound | OUTBound | HIGH
INBound
Measured power is within the range specified.
OUTBound
Measured power is out of the range defined.
HIGH
Measured power exceeds the upper scale limit.
*RST:
HIGH
Manual operation:
See
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit[:STATe]
<state>
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LIMit[:STATe]
<state>
Enables or disables the Out
1
/
Trig
Out connector as a monitoring output for the power
indication.
If enabled, you cannot use the connector for any other purpose.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
*RST:
OFF
Manual operation:
See
[SENSe<Sensor>:]POWer:REFLection:RANGe:LOWer
<lower>
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LOWer
<lower>
Effective if
[SENSe<Sensor>:]POWer:REFLection:RANGe:AUTO
Sets the lower scale limit for the power indication.
Measurement Settings and Results