Remote Control Commands
R&S
®
NRX
214
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer[:DATA]:POWer
<value>
Sets the lower limit for the measured power values.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-120.0 to +150.0
*RST:
-60.0
Default unit: dBm
Manual operation:
"Forward Lower Limit, Reflection Lower Limit"
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer:STATe
<state>
Enables or disables the monitoring function for the lower limit.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<state>
*RST:
OFF
Manual operation:
"Forward Lower Limit State, Reflection Lower Limit State"
CALCulate<Measurement>:LIMit<DirectionalChannel>:UPPer[:DATA]
<value>
Sets the upper limit for the measured values.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = forward, 2 = reflection (reverse)
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH