Firmware Update
R&S
®
NRX
171
User Manual 1178.5566.02 ─ 07
6. Check if the update was successful. This is the case if the firmware version in the
"Identification" field is the same as the one you loaded in the "Firmware" field.
12.2
Firmware Update via a USB Flash Memory Stick
This chapter contains information on installing/updating the firmware on the R&S NRX
via a USB flash memory stick.
12.2.1
Hardware and Software Requirements
The system requirements to perform a firmware update via a USB flash memory stick
are as follows:
●
PC or mobile device with free USB port running any operating system and software
that supports copying files to the USB flash memory stick
●
USB flash memory stick (USB 2.0 or 3.0, with USB-A plug, FAT32 file system, and
sufficient space for the firmware file)
●
A Rohde & Schwarz update file (*.rsu) for the sensor must be available.
12.2.2
Preparing an Update
To prepare an update via USB flash memory stick:
1. Copy the Rohde & Schwarz update file to the root directory of the USB flash mem-
ory stick.
2. Disconnect the USB flash memory stick from the PC or mobile device. If the instru-
ment is off, switch it on.
Firmware Update via a USB Flash Memory Stick