Remote Control Commands
R&S
®
NRP18S-xx
89
User Manual 1178.3686.02 ─ 05
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 8-7: Operation with 2-port device between signal source and sensor input
Configuring the S-Parameter Correction
[SENSe<Sensor>:]CORRection:SPDevice:LIST?
................................................................ 89
[SENSe<Sensor>:]CORRection:SPDevice:SELect
.............................................................. 89
[SENSe<Sensor>:]CORRection:SPDevice:STATe
............................................................... 89
[SENSe<Sensor>:]CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been downloaded to the power
sensor. The result of the query indicates the consecutive number and mnemonic of
each data set.
Usage:
Query only
[SENSe<Sensor>:]CORRection:SPDevice:SELect
<num>
Selects a downloaded data set for S-parameter correction.
See also
Chapter 8.8.4.3, "S-Parameter Correction"
Parameters:
<num>
Range:
1 to 1999
*RST:
1; can differ if a calibration set defines another
value.
Manual operation:
See
[SENSe<Sensor>:]CORRection:SPDevice:STATe
<state>
Activates or deactivates the S-parameter correction. If activated, uses the S-parameter
data set selected by
[SENSe<Sensor>:]CORRection:SPDevice:SELect
See also
Chapter 8.8.4.3, "S-Parameter Correction"
Parameters:
<state>
*RST:
OFF; can differ if a calibration set defines another
value.
Example:
CORR:SPD:SEL 1
Selects an S-parameter correction data set.
CORR:SPD:STAT ON
Activates the S-parameter correction.
Manual operation:
See
"S-Parameter Correction State Locked"
Configuring Basic Measurement Parameters