[SENSe<Sensor>:]TIMing:EXCLude:STARt
<exclude_start>
Sets a time that is to be excluded at the beginning of the integration period, see
Parameters:
<exclude_start>
Range:
0.0 to 1.0
*RST:
0.0
Default unit: Seconds
[SENSe<Sensor>:]TIMing:EXCLude:STOP
<exclude_stop>
Sets a time that is to be excluded at the end of the integration period, see
Parameters:
<exclude_stop>
Range:
0.0 to 1.0
*RST:
0.0
Default unit: Seconds
8.8.4
Configuring Corrections
It is possible to set some parameters that compensate for a change of the measured
signal due to fixed external influences.
........................................................................................... 86
.......................................................................................... 87
............................................................................................ 90
.................................................................................. 91
8.8.4.1
Duty Cycle Corrections
The duty cycle is the percentage of one period during which the signal is active, when
pulse-modulated signals are corrected. The duty cycle is only evaluated in the Continu-
ous Average mode.
Remote commands:
[SENSe<Sensor>:]CORRection:DCYCle
............................................................................ 86
[SENSe<Sensor>:]CORRection:DCYCle:STATe
.................................................................. 87
[SENSe<Sensor>:]CORRection:DCYCle
<duty_cycle>
Available in continuous average mode.
Sets the duty cycle for measuring pulse-modulated signals. The duty cycle defines the
percentage of one period during which the signal is active. If the duty cycle is enabled,
the R&S NRP18S-xx takes this percentage into account when calculating the signal
pulse power from the average power.