Concepts and features
R&S
®
ZNA
86
User Manual 1178.6462.02 ─ 20
4.1.3.1
Trace settings
The trace settings specify the mathematical operations used to obtain traces from the
measured or stored data. They can be divided into several main groups:
●
Selection of the measured quantity (S-parameters, wave quantities, ratios, impe-
dances,...)
●
Conversion into the appropriate display format and selection of the diagram type
●
Scaling of the diagram and selection of the traces associated to the same channel
●
Readout and search of particular values on the trace by means of markers
●
Limit check
The trace settings can be accessed via the keys in the Trace section of the (virtual) key
panel. They complement the
accessible via the Stimulus and Chan-
nel sections.
Each trace is assigned to a channel. The channel settings apply to all traces of the
channel.
4.1.3.2
Channel settings
A channel contains hardware-related settings which specify how the network analyzer
collects data. The channel settings can be divided into three main groups:
●
Description of the test setup (power of the internal source, IF filter bandwidth, port
configuration, receiver step attenuators, ...)
●
Control of the measurement process (sweep, trigger, averaging, ...)
●
Correction data (calibration, offset, ...)
The channel settings can be accessed via the Stimulus and Channel sections of the
(virtual) key panel.
Basic concepts