GUI reference
R&S
®
ZNA
538
User Manual 1178.6462.02 ─ 20
Cal. Power Settings tab
The "Cal Power Setting" tab collects all calibration-related power settlings that were
previously distributed across several softtool tabs and dialogs.
Source Power
See
Source Step Att
Source step attenuator setting used during calibration, if activated.
Remote command:
[SENSe<Ch>:]CORRection:ADVanced:SOURce<PhyPt>:ATTenuator
[SENSe<Ch>:]CORRection:ADVanced:SOURce<PhyPt>:ATTenuator:STATe
ZNA Port Power
"Power Result", as displayed in the
Cal Power Offset
For power calibrations only: Specifies a gain (positive values) or an attenuation (nega-
tive values) in the signal path between the source port and the calibrated reference
plane. With a "Cal Power Offset" of n
dB, the target power at the reference plane (cal
power) is equal to the actual output power at the port plus n
dB. The "Cal Power Off-
set" has no impact on the source power.