GUI reference
R&S
®
ZNA
509
User Manual 1178.6462.02 ─ 20
... /Logic/ ...
The table defines all settings related to "Multiple Triggers". For all measurement
sequences it is possible to select (and configure) the related trigger logic:
●
no trigger required ("Free Run")
●
a single trigger source with configurable
●
a logical combination of two trigger sources, each with configurable "Signal Type"
Remote command:
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce<inp>
TRIGger<Ch>[:SEQuence]:MULTiple:LOGic
TRIGger<Ch>[:SEQuence]:MULTiple:SLOPe<Num>
TRIGger<Ch>[:SEQuence]:MULTiple:HOLDoff
5.10.3.3
Trigger Out Manager dialog
Defines if and how the channel-specific sweep sequence is indicated at the Trigger Out
interfaces.
Access:
Channel – [Sweep] > "Trigger" > "Trigger Out Manager..."
Trigger Out connectors A to D are provided by the optional
ware option is not installed, it is not possible to generate outgoing trigger signals. In
particular, the "Trigger Out Manager" dialog is not available in this case.
Sequence
Selects the sequence to be indicated at the related Trigger Out "Connector". Trigger
out signals can be generated for every sweep, segment, partial measurement, and
sweep point.
Furthermore, trigger out signals can be generated for the data acquisition phases and
for the start of pulse generator pulses.
Remote command:
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
Signal Type
Defines how the selected
is indicated at the related Trigger Out "Connector":
●
using pulses with positive or negative polarity ("Pulse Positive", "Pulse Negative")
Sweep Softtool