GUI reference
R&S
®
ZNA
312
User Manual 1178.6462.02 ─ 20
●
For a balanced port, two physical ports are combined to form a single logical port.
●
For unused ports, the physical port is crossed out; no logical port number is
assigned.
Select Predefined Port
Allows you to select from a set of predefined logical port configurations.
The available configurations depend on the number of analyzer ports.
Remote command:
SOURce<Ch>:LPORt<LogPt>
SOURce<Ch>:LPORt<LogPt>:CLEar
True Differential Mode
Activates/deactivates True Differential Mode
This checkbox is only visible if
option R&S
ZNA-K61 is available.
It is enabled, if at least one balanced port is configured, whose physical ports are fed
by independent sources (see
Chapter 4.7.19, "Internal 3rd and 4th source for 4-port
on page 273). In case the balanced port is made up of frequency converter
ports,
must be active for both of them.
Use the "Source Cohrence..." button to open the
dialog,
which allows you to configure the TruDi-specific sweep types.
Remote command:
Source Coherence...
Opens the dialog of the same name, which allows you to configure and enable source
coherence and to configure the true differential mode sweep types "Phase Imbalance"
and "Amplitude Imbalance".
See
Chapter 5.13.3.2, "Source Coherence dialog"
User Config tab
The "User Config" tab of the "Balanced Ports" dialog defines a new balanced port con-
figuration.