Command reference
R&S
®
ZNA
1489
User Manual 1178.6462.02 ─ 20
Example:
// Select the ICS
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess 'ZN-Z30-02::100010'
// Query for the connected ICUs
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:SUBModule:ALL?
// '[1]ZN-Z33-02::101325,[2]ZN-Z33-02::101336,[3]ZN-Z33-02::101339'
// Activate ICU[1]
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess 'ZN-Z33-02::101336'
// Query the available characterizations
:SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
// 'Factory,afterAwhile'
// Select the non-factory characterization
:SYSTem:COMMunicate:RDEVice:AKAL:CKIT 'afterAwhile'
// Query the active characterization
:SYSTem:COMMunicate:RDEVice:AKAL:CKIT?
// 'ZN-Z33-02::101336=afterAwhile'
Manual operation:
See
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
Queries all characterizations (cal kit files) which are stored on the connected calibra-
tion unit, either on the calibration unit's internal (flash) memory or on an SD card inser-
ted at the calibration unit (prefix: "SD"). A possible response is
'Factory,ZN-Z51_custom,Throughs,SD:test'
. The factory characterization is
always available; an empty string denotes that no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess