Command reference
R&S
®
ZNA
1487
User Manual 1178.6462.02 ─ 20
The receive terminator setting is relevant if block data is transferred to the analyzer
(
REAL
). In the default setting LFEoi, the analyzer recognizes an LF
character sequence with or without the EOI control bus message as a receive termina-
tor. An accidental LF in a data block can be recognized as a terminator and cause an
interruption of the data transfer.
The EOI setting is especially important if commands are transferred in block data for-
mat, because it ensures that the parser for command decoding is activated by the ter-
minator only after the command has been completely transferred. Readout of binary
data does not require a change of the receive terminator.
Parameters:
<Terminator>
LFEoi | EOI
LFEoi
A line feed character sequence with or without EOI is recognized
as receive terminator
EOI
Only EOI is recognized as receive terminator
Example:
SYST:COMM:GPIB:RTER EOI
Set the terminator to EOI.
SYSTem:COMMunicate:NET:HOSTname
<HostName>
Sets or gets the host name of the instrument.
Parameters:
<HostName>
Host name
*RST:
A preset of the instrument doesn't change the host
name.
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
<Address>
Selects one of the USB-connected calibration units for calibration (see commands
SENSe<Ch>:CORRection:COLLect:AUTO...
). This command is not necessary if
only one cal unit is connected.
Parameters:
<Address>
Name (USB address) of a connected calibration unit (string vari-
able). The names of all connected cal units can be queried using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
Example:
SYST:COMM:RDEV:AKAL:ADDR:ALL?
Query the names of all connected calibration units.
SYST:COMM:RDEV:AKAL:ADDR 'ZV-Z52::1234'
Select the cal unit named 'ZV-Z52::1234' for calibration.
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port TOSM calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Manual operation:
See
SCPI command reference