Command reference
R&S
®
ZNA
1395
User Manual 1178.6462.02 ─ 20
7.3.14.26
[SENSe:]SPECtrum...
............................................................................... 1395
[SENSe<Ch>:]SPECtrum:IREJection
............................................................................. 1395
................................................................................ 1396
[SENSe<Ch>:]SPECtrum:DETector
<Type>
Defines the detector for the spectrum measurement, i.e. how the measurement result
for a particular sweep point is derived from the related sample values.
Suffix:
<Ch>
.
Parameters:
<Type>
MAXPeak | MINPeak | RMS | AVERage
MAXPeak
Take the largest of all sample values
MINPeak
Take the smallest of all sample values
RMS
Take the root mean square of all sample values
AVERage
Take the arithmetic mean of all sample values
Options:
R&S
ZNA-K1
Manual operation:
See
[SENSe<Ch>:]SPECtrum:IREJection
<Mode>
Selects the image rejection strategy: the higher the number of data acquisitions, the
better the image rejection.
Suffix:
<Ch>
.
Channel number
Parameters:
<Mode>
OFF | LOW | NORMal | HIGH
OFF
1 data acquisition (no image rejection)
Recommended if you want to evaluate the noise power ratio
(
CALCulate<Chn>:STATistics:PRATio[:STATe] ON
LOW
2 acquisistions
NORMal
4 acquisistions
HIGH
8 acquisistions
*RST:
NORMal
SCPI command reference