Command reference
R&S
®
ZNA
1358
User Manual 1178.6462.02 ─ 20
FB
The IF frequency at IF Reference <Pt> and IF Meas <Pt>
sweeps with the channel base frequency.
CW
Fixed IF frequency at IF Reference <Pt> and IF Meas <Pt>.
*RST:
SWEep
Usage:
Setting only
Options:
R&S
ZNA-B26
Manual operation:
See
[SENSe<Ch>:]PATH<Pt>:IFSWitch
<Selection of IF Switch on Ref and Meas path>
Defines if and how the Direct IF Access Input / Output ports IF Reference <Pt> and IF
Meas <Pt> are used in channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Setting parameters:
<Selection of IF
Switch on Ref and
Meas path>
OFF | IN | OUT
OFF
The ports are unused.
IN
The ports are for IF input (but not for IF output).
Note that direct generator/receiver access cannot be used in this
case.
OUT
The ports are for IF output (but not for IF input).
Usage:
Setting only
Options:
R&S
ZNA-B26
Manual operation:
See
[SENSe<Ch>:]PATH<Pt>:MEASurement:DIRectaccess?
[SENSe<Ch>:]PATH<Pt>:REFerence:DIRectaccess?
Queries the
direct generator/receiver access
path.
To define the direct generator/receiver access path, use
.
Note that direct generator/receiver access and direct IF input cannot be used at the
same time. Direct generator/receiver access is only used if
is set to
OFF
or
OUT
.
Suffix:
<Ch>
.
Channel number
SCPI command reference