Command reference
R&S
®
ZNA
1289
User Manual 1178.6462.02 ─ 20
Example:
CORR:COLL:AUTO:TYPE FNPort, '', 1, 2, 4
Perform an automatic full 3-port calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:UTHRough
<TestPort1>, <TestPort2>
With this command, you can perform the Unknown Through measurement during an
automatic UOSM calibration.
Although it works for all calibration units, it is only useful if the calibration unit selected
using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
(ICS).
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TestPort1>,
<TestPort2>
VNA test port numbers
Example:
See
Chapter 8.2.5.5, "Inline calibration"
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect:CKIT:INSTall
<CalKitFile>, <Gender>[,
<Ports>]
Allows you to load cal kit data by gender:
●
From the given file
●
To the given ports or all ports
Use
on page 1181 to get information about cal kit files.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<CalKitFile>
Path to the cal kit file, either absolute or relative to the current
directory (see
<Gender>
MALE | FEMale | NGENder
<Ports>
Either a comma-separated list of port numbers or ALL (optional)
Usage:
Setting only
SCPI command reference