User Manual 1179.4363.02 ─ 01
Example:
//Select the SNMP version
SYST:COMM:SNMP:VERS V12
11.9.7
Checking the System Configuration
Useful commands to check the system configuration described elsewhere:
●
.............................................................................................503
...................................................................................... 504
........................................................................................ 504
..................................................................................... 504
............................................................................................ 505
........................................................................................... 505
................................................................................................506
............................................................................................506
DIAGnostic:INFO:CCOunt?
<Relay>
This command queries how many switching cycles the individual relays have per-
formed since they were installed.
Query parameters:
<Relay>
ACDC
Mechanical Attenuation Coupling
ATT5
Mechanical Attenuation 05 DB
ATT10
Mechanical Attenuation 10 DB
ATT20
Mechanical Attenuation 20 DB
ATT40
Mechanical Attenuation 40 DB
CAL
Mechanical Calibration Source
EATT
Electrical Attenuation Bypass
PREamp
Preamplifier Bypass
Return values:
<Cycles>
Number of switching cycles.
Example:
DIAG:INFO:CCO? CAL
Usage:
Query only
Configuring the R&S
FSPN