Remote Control
R&S
®
FSPN
486
User Manual 1179.4363.02 ─ 01
SOURce<si>:TEMPerature:FRONtend
This command queries the current frontend temperature of the R&S
FSPN.
During self-alignment, the instrument's (frontend) temperature is also measured (as
soon as the instrument has warmed up completely). This temperature is used as a ref-
erence for a continuous temperature check during operation. If the current temperature
deviates from the stored self-alignment temperature by a certain degree, a warning is
displayed in the status bar indicating the resulting deviation in the measured power lev-
els. A status bit in the
STATUs:QUEStionable:TEMPerature
register indicates a
possible deviation.
Suffix:
<si>
.
irrelevant
Return values:
<Temperature>
Temperature in degrees Celsius.
Example:
SOUR:TEMP:FRON?
Queries the temperature of the frontend sensor.
11.9.4
Customizing the Screen Layout
............................................................................... 486
.............................................................................489
...................................................................................... 491
11.9.4.1
Configuring Screen Elements
Useful commands to configure screen elements described elsewhere:
●
●
............................................................................................. 486
.....................................................................................487
....................................................................................... 488
...........................................................................................488
DISPlay[:WINDow<n>]:TIME:FORMat
..............................................................................488
................................................................................... 489
DISPlay:ANNotation:CBAR
<State>
This command hides or displays the channel bar information.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Configuring the R&S
FSPN