Blinking orange Data transfer between the calibration unit and the network analyzer. Do not

disconnect the USB cable.

Blinking blue

Calibration in progress. Do not disconnect the USB cable.

4.3

 

Performing an automatic calibration

Once the USB connection between calibration unit and network analyzer has
been established, an automatic calibration can be performed as follows:

1. Start the automatic calibration function of your network analyzer.

2. Configure the automatic calibration. In particular:

a) Select the test ports to calibrate, and the desired calibration type.
b) Select the appropriate R&S

 

ZN-Z15x model and characterization.

c) Configure the required RF port assignments between analyzer and calibra-

tion unit.

3. For each RF port assignment:

a) Connect the calibration unit ports to the test ports of the network analyzer,

as depicted in the analyzer GUI.

b) Start the calibration data acquisition.

4. Disconnect the RF ports of the calibration unit.

Wear a grounded wrist strap during steps 3 and 4 (see 

"Preventing electro-

static discharge (ESD)"

 

on page 7).

Accuracy considerations

To ensure an accurate calibration, please observe the following items:

Do not use adaptors between the calibration unit and the test ports.

After connecting the unit to the USB port, allow for a sufficient warm-up time
(see data sheet) before starting the calibration.

To ensure best accuracy, the analyzer automatically reduces the source power
to –10 dBm. If the test setup contains a large attenuation, deactivate Auto-
matic Power Reduction for Calibration Unit in the Calibration tab of the Sys-
tem Configuration dialog and ensure an input power of –10 dBm at the ports
of the calibration unit (please also refer to the data sheet).

Performing an automatic calibration

Содержание 1317.9134.32

Страница 1: ...R S ZN Z15x Calibration Units User Manual 1337596302 Version 04 U 2...

Страница 2: ...f 2 ports 5 GHz to 67 GHz 1332 7239 02 R S ZN Z156 1 85 mm f 2 ports 10 MHz to 67 GHz 1332 7239 03 User manual 1326 5080 02 covers the R S ZN Z154 calibration unit 2022 Rohde Schwarz GmbH Co KG Muehl...

Страница 3: ...roduction and verification of electronic components and devices in industrial administrative and laboratory environments Use the R S ZN Z15x only for its designated purpose Observe the operating condi...

Страница 4: ...ning messages in the documentation A warning message points out a risk or danger that you need to be aware of The signal word indicates the severity of the safety hazard and how likely it will occur i...

Страница 5: ...do not wear out because they are switched electroni cally The R S ZN Z15x calibration units differ in the number of calibration ports and the connector types For an overview refer to the data sheet 3...

Страница 6: ...esidential buildings Class A equipment is intended for use in industrial environments It can cause radio disturbances in residential environments due to possible conducted and radiated disturbances It...

Страница 7: ...nts of the product and the device under test DUT Ground yourself to prevent electrostatic discharge damage a Use a wrist strap and cord to connect yourself to ground b Use a conductive floor mat and h...

Страница 8: ...See also Cable selection and electromagnetic interference EMI on page 6 3 5 RF connections The calibration unit provides two RF connectors numbered 1 to 2 to be connec ted to the test ports of the ana...

Страница 9: ...ialization the status LED will switch to green see Chapter 4 2 Function of the status LED on page 9 The unit is ready to use as outlined below 4 2 Function of the status LED The LED on top of the Cali...

Страница 10: ...ssignment a Connect the calibration unit ports to the test ports of the network analyzer as depicted in the analyzer GUI b Start the calibration data acquisition 4 Disconnect the RF ports of the calib...

Страница 11: ...of your network analyzer for this purpose For details refer to the analyzer s help system 4 5 microSD card On the rear of the R S ZN Z152 and R S ZN Z153 calibration units there is a microSD card slot...

Страница 12: ...ct contact our customer support center A team of highly qualified engineers provides support and works with you to find a solution to your query on any aspect of the operation program ming or applicat...

Отзывы: