Command reference
R&S
®
ZNB/ZNBT
845
User Manual 1173.9163.02 ─ 62
CALCulate:CALValidate:CHARacteriza
<String>
Selects the CalU characterization to be used for the cal validation.
Use
CALCulate:CALValidate:STANdard
to select the CalU standard to be used
for the validation.
Parameters:
<String>
*RST:
'Factory'
Manual operation:
See
CALCulate:CALValidate:RESPonse:MAGNitude:LOWer
<Port>, <Value>
CALCulate:CALValidate:RESPonse:MAGNitude:UPPer
<Port>, <Value>
Defines the maximum magnitude deviations from the reference characterization S-
parameters that are allowed during cal validation.
Use
CALCulate:CALValidate:CHARacteriza
to select the reference characteri-
zation.
Parameters:
<Port>
Physical analyzer port no.
<Value>
*RST:
-0.2 (LOWer) | +0.2 (UPPer)
Default unit: dB
Manual operation:
See
CALCulate:CALValidate:RESPonse:PHASe:LOWer
<Port>, <Value>
CALCulate:CALValidate:RESPonse:PHASe:UPPer
<Port>, <Value>
Defines the maximum phase deviations from the reference characterization S-parame-
ters that are allowed during cal validation.
Use
CALCulate:CALValidate:CHARacteriza
to select the reference characteri-
zation.
Parameters:
<Port>
<Value>
*RST:
-2 (LOWer) | +2 (UPPer)
Default unit: degrees
Manual operation:
See
CALCulate:CALValidate:RUN
Runs the cal validation with its configured settings.
Usage:
Event
Manual operation:
See
SCPI command reference