GUI reference
R&S
®
ZNB/ZNBT
624
User Manual 1173.9163.02 ─ 62
Figure 5-58: Offset Embed > Single Ended softtool
left = 2 port data file selected for deembedding
right = lumped element model selected for de-/embedding
If the "Fixture Simulator" is disabled for the related channel (see
on page 549), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
Overview
See
Type
Switches between "Deembedding" and "Embedding" network definition.
Port
Physical port. The transformation networks are defined such that the analyzer is con-
nected to the left of the circuit while the DUT is connected to the right side. You can
define independent transformation networks for all ports.
Remote command:
The
<PhyPt>
numeric suffix in the embedding/deembedding commands identifies the
CALCulate<Ch>:TRANsform:VNETworks:SENDed:
or
CALCulate<Ch>:TRANsform:VNETworks:
SENDed:EMBedding<PhyPt>[:STATe]
.
All Deembedding Activated
Deactivates or activates all deembeddings (single-ended, balanced, port set, ground
loop) for the active channel, if any.
This control is only visible if
Offset Embed softtool