Command reference
R&S
®
ZNB/ZNBT
1393
User Manual 1173.9163.02 ─ 62
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
<CalKitName>, <Type>,
<SParameter>[, <FirstPort>, <SecondPort>]
Reads the calibration data for a particular standard from a cal unit characterization
(calkit file). If more than one calibration unit is connected, the related one must be
selected using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
.
Query parameters:
<CalKitName>
String parameter containing the name of a cal unit characteriza-
tion (calkit file).
'Factory'
denotes the factory characterization; an empty
string
' '
denotes the last referenced characterization; the pre-
fix "SD:" indicates that the characterization is stored on an SD
card inserted at the calibration unit.
<Type>
THRough | OPEN | SHORt | MATCh | MMTHrough |
MFTHrough | FFTHrough | MOPen | FOPen | MSHort | FSHort |
MMTCh | FMTCh
Standard types; for a description refer to table
. The standard types of a particular charac-
terization can be queried via
RDEVice:AKAL:CKIT:STANdard:CATalog?
The factory characterization usually does not contain data for a
through standard; therefore a query of the type
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
'Factory', THR, S11, 1, 2
results in an error message.
<SParameter>
S11 | S12 | S21 | S22
S-parameter of the standard, use
S11
for one-port standards,
one of the four for 2-port standards.
<FirstPort>
First port number (sufficient for one-port standards). Port num-
bers can be omitted if the cal kit data is valid for all ports; see
<SecondPort>
Second port number, for two-port standards. Port numbers can
be omitted if the cal kit data is valid for all ports.
Example:
See
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Queries the warmup state of the connected calibration unit R&S ZV_Z5x. If several cal
units are USB-connected to the analyzer, the command queries the cal unit selected
via
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
SCPI command reference