Command reference
R&S
®
ZNB/ZNBT
1287
User Manual 1173.9163.02 ─ 62
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
"Permittivity / Velocity Factor"
[SENSe<Ch>:]HARMonic:RTIMe:DATA
<DUTRiseTime>
Sets/gets the minimum rise time the user wishes to measure on the DUT.
Use
[SENSe<Ch>:]HARMonic:RTIMe:THReshold
to select the underlying rise time
definition.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets
to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTRiseTime>
Default unit: s
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
[SENSe<Ch>:]HARMonic:RTIMe:THReshold
<DUTRiseThreshold>
Defines how the rise time that is set using
[SENSe<Ch>:]HARMonic:RTIMe:DATA
shall be interpreted. Allows to select between rise time definitions 20%-80% (default)
and 10%-90%.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets
to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTRiseThreshold> T1_9 | T2_8
Rise time definition:
T1_9
: 10%-90%
T2_8
: 20%-80%
*RST:
T2_8
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
[SENSe<Ch>:]HARMonic:VELocity:DATA
<DUTVelocity>
Sets/gets the velocoty factor of the DUT.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets
to ON.
SCPI command reference