Basic Operation

R&S

®

 ZVA-Z75, -Z90, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500

22

Quick Start Guide 1307.7039.62 ─ 10

After the power calibration procedure a system error correction is recommended. Due
to the physical properties of the mm-waves and the waveguides, measurements with
frequency converters require a special calibration kit for system error correction.

Rohde & Schwarz offers kits for this purpose:

Calibration kit R&S

 

ZV-WR15, for converter R&S

 

ZVA-Z75

Calibration kit R&S

 

ZV-WR12, for converter R&S

 

ZVA-Z90

Calibration kit R&S

 

ZV-WR10, for converter R&S

 

ZVA-Z110

Calibration kit R&S

 

ZV-WR08, for converter R&S

 

ZVA-Z140

Calibration kit R&S

 

ZV-WR06, for converter R&S

 

ZVA-Z170

Calibration kit R&S

 

ZV-WR05, for converter R&S

 

ZVA-Z220

Calibration kit R&S

 

ZV-WR03, for converter R&S

 

ZVA-Z325

Calibration kit R&S

 

ZV-WR02, for converter R&S

 

ZVA-Z500

The standards in the calibration kit allow all one-port and two-port calibration types
supported by the network analyzer except TNA. Refer to the documentation of the cali-
bration kit or the help system of your network analyzer for details.

3.6 Measurement

After power calibration and system error correction, the mm-wave measurement can
be performed like any other network analyzer measurement. The "Port Configuration"
settings (together with the "Stimulus" settings), determine the sweep range of the con-
verted signals (i.e. the input and output frequencies at the DUT ports). All measured
quantities (S-parameters, wave quantities, ratios etc.) and other trace settings are
available.

The following restrictions hold for measurements with external frequency converters:

Power flatness calibration and power sweep are only available for the R&S

 

ZVA-

Z90 and the latest variant of the R&S

 

ZVA-Z110 (order number 1307.7000.0

3

); for

these converter types, use the "Frequency Converter Leveling Tool" to linearize the
output power in the frequency range of interest.

In order to adjust the output power of other converter models (e.g. for measuring
wave quantities or testing compression effects), use the adjusting screw on top of
the converters (see 

chapter 2.2, "Output Power Adjusting Screw"

on page 7). This

is also possible with the R&S

 

ZVA-Z110.

The following example shows the transmission and reflection coefficients of a band-
pass filter in the frequency range between 75 GHz and 110 GHz, which is covered by
the frequency converter R&S ZVA-Z110.

 

Measurement

Содержание 1307.7000.03

Страница 1: ...R S ZVA Z75 Z90 Z110 Z140 Z170 Z220 Z325 Z500 Converters Quick Start Guide Quick Start Guide 1307 7039 62 10 7 W Test Measurement...

Страница 2: ...220 Converter WR05 stock no 1307 8006 02 R S ZVA Z325 Converter WR03 stock no 1317 0514 02 R S ZVA Z500 Converter WR02 stock no 1317 0520 02 2015 Rohde Schwarz GmbH Co KG M hldorfstr 15 81671 M nchen...

Страница 3: ...2 2 Output Power Adjusting Screw 7 2 3 Rear Panel 8 2 4 Putting the Converter into Operation 10 2 5 Maintenance 17 2 6 Storing and Packing 17 3 Basic Operation 18 3 1 Required Equipment 18 3 2 Measur...

Страница 4: ...Contents R S ZVA Z75 Z90 Z110 Z140 Z170 Z220 Z325 Z500 4 Quick Start Guide 1307 7039 62 10...

Страница 5: ...rce power of the network analyzer The frequency converter mode ensures compat ible source powers Before you connect your converter to the network analyzer always activate the fre quency converter mode...

Страница 6: ...the instrument R S ZVA Z500 the other models are similar in design The precision waveguide flange of the test port adapter is equipped with two alignment pins and two holes receiving the alignment pin...

Страница 7: ...djusting Screw The output power of the converter can be adjusted via the network analyzer only for the models ZVA Z90 and ZVA Z110 The converter models R S ZVA Z75 R S ZVA Z140 R S ZVA Z170 R S ZVA Z2...

Страница 8: ...er RF Connectors Input RF Connectors Output Fig 2 3 Rear view of the frequency converter 2 3 1 Standby Switch The standby toggle switch connects ready state or disconnects standby state the internal m...

Страница 9: ...t voltage and current must not exceed the maximum values according to the rear panel labeling or the data sheet Notice that the external DC power supply units of R S ZVA Z500 converters support higher...

Страница 10: ...rs Output Two SMA connectors MEAS OUT measurement signal output REF OUT reference signal output For correct cabling please refer to chapter 2 4 5 Connecting RF Cables on page 13 2 4 Putting the Conver...

Страница 11: ...tach one of the included protective caps to the adapter Also attach protective caps to the second test port adapter The wooden box should also be kept and used for storage of the instrument and the ac...

Страница 12: ...vailable see chapter 2 1 Test Port Adapter Waveguide Flange on page 6 Select the appropri ate adapter depending on the DUTs to be measured Standard adapter R S ZVA Z110 R S ZVA Z140 R S ZVA Z170 R S Z...

Страница 13: ...sible These are mechanically compatible with 3 5 mm con nectors while providing tolerable electrical mismatch e g R S ZV Z195 order number 1306 4536 xx For a complete test setup for a 2 port transmiss...

Страница 14: ...m to 2 92 mm adapters to con nect the cables These adapters are offered as complementary adaption kits see Adaption Kits R S ZCAK on page 14 1 Connect MEAS OUT of the converter to the NWA Use the MEAS...

Страница 15: ...d adapter press the small button next to the adapter and push the adapter away from the button Connect the power supply to the 9 V MAX 1 1 A DC input on the rear panel see chapter 2 3 2 Power Supply C...

Страница 16: ...ssory A tight and accurate connection is very important to ensure precise calibration and measurement results Depending on the type of waveguide flange of the DUT it may be required to exchange the te...

Страница 17: ...eling or plastic parts Never use cleaning agents such as solvents thinners acetone etc acids bases or other substances For our support center address and a list of useful R S contact addresses refer t...

Страница 18: ...A Z220 at least version V2 72 R S ZVA Z325 at least version V2 45 R S ZVA Z500 at least version V2 80 Two NWA ports per Frequency Converter port Alternatively One NWA port per Frequency Converter plus...

Страница 19: ...rter Type R S ZVA Z75 R S ZVA Z90 R S ZVA Z110 R S ZVA Z140 R S ZVA Z170 R S ZVA Z220 R S ZVA Z325 or R S ZVA Z500 3 Select a test setup with an analyzer port as external source click Apply to acti va...

Страница 20: ...er calibration chapter 2 4 8 Mounting a DUT on page 16 3 5 Calibration The R S ZVA Z90 and the latest variant of the R S ZVA Z110 order number 1307 7000 03 allow for a variation of the output power by...

Страница 21: ...ion over frequency and power is needed use the R S ZVA Frequency Converter Levelling Tool Free download from the R S ZVA Z Millimeter Wave Converters site A receiver power calibration of the reference...

Страница 22: ...ment After power calibration and system error correction the mm wave measurement can be performed like any other network analyzer measurement The Port Configuration settings together with the Stimulus...

Страница 23: ...e including remote control refer to the R S ZVA ZVT online help system or to the printable operating man ual which is available for download at http www rohde schwarz com product zva Application notes...

Страница 24: ...9 M Maintenance 17 MEAS OUT 10 Measurement 22 Measurement principle 18 P Power adjusting screw 7 Power calibration 20 Power on 15 Power supply connector 9 Preparing for use 6 R Rear panel 8 REF OUT 10...

Отзывы: