Technical Note TN-136 

06/10/TM

RAE Systems by Honeywell  877-723-2878  raesystems.com

3

Once the tube is in place, begin measuring by pressing [Y/+]. 
The display shows a countdown (60 seconds is shown here, but 
sampling time depends on the type of separation tube selected and 
the temperature):

Note:

 You can abort the sampling by pressing [N/-] at any time.

Once the countdown is complete, the reading is shown:

Press [Y/+] to continue sampling with the tube for 15 minutes to 
establish a STEL reading, or press [N/-] to return to the main menu.

WARNING!

 At least 1/4 of the tube should still be yellow-

orange at the bottom. If not, the STEL value is not valid. Abort the 
measurement and change the tube. Then do a snapshot test instead 
of a STEL test. 

Note:

 If the STEL is exceeded, the UltraRAE 3000 

goes into alarm.

If you press [N/-] to return to the main menu, which shows the tube 
type instead of the CF (correction factor):

Press [N/-] to advance to this screen:

If you press [Y/+], you are asked, “Clear peak value! Are You Sure?” 
to confirm:

Press [Y/+] to clear the Peak value and exit to VOC operation.

If you press [N/-], this display is shown:

Remove the tube and put the inlet back together. Then press [N/-]. 
This display is shown:

After a few seconds, the UltraRAE 3000 enters VOC mode and 
shows this display:

You can step through the rest of the steps by pressing [N/-] 
repeatedly until you reach the main menu again.