- 72 -
20. ERROR COMMANDS
Err2
- Value beyond the zero range
Err3
- Value beyond the tare range
Err4
- Calibration mass or start mass beyond the acceptable
range (
±
1% for weight,
±
10 for start mass)
Err5
- Mass of a single piece lower than the scale division
Err8
- Exceeded the time for tarring, zeroing, start mass
adjustment or span adjustment
Err9
- Time for internal weight lifting/dropping down exceeded
(refers to WLC…/C/2 scales)
NULL
- Zero value from the AD converter
FULL2
- Measurement range
overflow
LH
- Start mass error, the mass on the weighing platform is
beyond the acceptable range (-5% to +15% of start mass)
5-FULL
- Display range overflow in totalizing
Notice:
1. Errors: Err2, Err3, Err4, Err5, Err8, Err9, null, that appear on the
display are also signalled by a short beep sound (about 1 sec.);
2. Error FULL2 that appears on the display is also signalled by
a continuous sound until the cause of error disappears.
21. TROUBLE SHOOTING
Problem
Cause
Solution
Turning on does not
work
Discharged batteries.
Connect to mains or
change batteries
No batteries (not installed or
improperly installed)
Check the correctness of
installation (polarization)
The scale turns off
automatically
„t1” set to „YES” (Power
save)
In „othr” submenu change
„5.4 t1” to „no”
After turning on „LH”
message on the display
Loaded weight pan during
powering up
Unload the pan. Then the
scale will indicator zero.
Содержание WPT/4
Страница 2: ...2 AUGUST 2016...
Страница 10: ...10 4 MAIN DIMENSIONS 4 1 Low profiled scales...
Страница 11: ...11 4 2 Ramp scales 4 3 Stainless steel scales...
Страница 12: ...12 4 4 Stainless steel scales pit version 4 5 Mild steel scales...
Страница 13: ...13 4 6 Pallet scales 4 7 Beam scales...
Страница 32: ...32 Return to weighing see 10 2 2 13 3 Baud rate Procedure Enter the submenu P2 Prnt and then...
Страница 74: ...74...
Страница 75: ...75...