78
26.4.2. Continuous transmission
Results are transmitted in basic units. Transmission can be disabled/enabled
using RS232 command:
C1 CR LF
– Switch on continuous transmission in basic unit
C0 CR LF
– Switch off continuous transmission in basic unit
or using the parameter:
settings/ RS-232/Aut. printout/continuous
The frame length is constant (21 characters).
Frame format:
[SI] [space] [stability] [space] [sign] [mass] [space] [unit] [CR LF]
Stability character – [space] if stable
[?] if not stable
[^] if an indication over the range
[v] if an indication below the range
sign – [space] for positive values or [-] for negative values
mass – 9 characters justified to the right
unit – 3 characters justified to the left
26.4.3. Continuous transmission „with pauses”
Continuous transmission „with pauses” allows users to determine the time
interval between the subsequent frames in the parameter
interval.
His type of printouts can be enabled only in parameters:
Settings/RS-232/ Aut.
printouts/with interv.
The frame format is the same as in basic simple transmission.
The interval is set in:
Setting/RS-232/Interval
26.4.4. Stable result transmission
The scale automatically generates after each stable result separated by a
stable result below the
Min. mass
that can be set in parameters.
The intended use of this printout is for slip printers so it doesn’t have a starting
characters and the length can vary from 18 to 20 characters.