33
-
process of calibration can be postponed frequently, but fact that long postponing
of calibration can be cause of bigger error during weighing should be noticed.
These errors are effect of changes of temperatures and as consequence
changes of balance sensitivity.
Automatic system includes 3 possibilities:
-
calibration with regard to changes of temperature
balance is provided with precise system of controlling temperature
changes, temperature of calibration is registered every time, next is started
if temperature of surrounding changes 3
o
C
-
calibration with regard to passage of time
user can declare 4 spaces of time that makes criterion for calibration,
following option are accessible: calibration every 1 - 12 hours.
11.2. Entering calibration menu
P1 CAL
P1.1 iCAL
|
P1.2 uCAL
|
P1.3 tCAL
|
P1.4 ACAL
|
both
P1.5 CALt
|
1
h
P1.6 CALr
|
YES