- 92 -
available
units
Response:
UI_”g, mg, ct”_OK CR LF
– response:
available
units
Set current unit
Command overview: Command sets current unit for a particular device.
Format: US_x CR LF
Response options:
US_ x_OK CR LF
– command carried out, response:
current unit
US_E CR LF
– error in-course of command
execution, no parameter or
incorrect format
US_I CR LF
– command understood but not
accessible at this moment
x
– parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, msg,
u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates „
” button
pressing).
An example:
Command:
US_mg CR LF
– set „mg”
unit
Response:
US_mg_OK CR LF
– „mg” set as
a current unit
Give current unit
Command overview: command returns current unit.
Format: UG CR LF
Response options:
UG_x_OK CR LF
– command carried out, response:
current unit
UG_I CR LF
–
command understood but not
accessible at this moment
x
– parameter, unit symbol
An example:
Command:
UG CR LF
– return current
unit
Response:
UG_ct_OK CR LF
– currently set unit
is „ct”
Activate sound signal
Command overview: command activates BEEP sound signal for a specified amount of time
Format: BP_CZAS CR LF
Response options:
BP_OK CR LF
– command carried out, BEEP sound
signal activated
BP_E” CR LF
– no parameter or incorrect format
BP_I CR LF
–
command understood but not
accessible at this moment
TIME
– parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 ÷ 5000>.
If value greater than the permissible high limit is given, than BEEP sound is operated for the
maximum permissible amount of time.
An example:
Command:
BP_350 CR LF
– activate BEEP for
350 ms
Response:
BP_OK CR LF – BEEP activated
Содержание PS R1
Страница 99: ...99...