CALIBRATION TEST – adjustment process test
Fuction designed to enable comparison of intenal adjustment weight with its value
saved in scale memory. The process is carried out automatically, its result is
displayed on the screen (if the scale is connected via communication port with
a computer or a printer, the adjustement test result is printed out). Press
ESC
to go
back to the previous menu.
AUTO.CALIB. – automatic adjustment (for scale with internal adjustment)
Specyfying factor that is to trriger the automatic internal adjustment
NONE
neither time nor temperature (automatic adjustment turned off),
TEMP.
change of scale inside temperature,
TIME
time interval set in
P1.6 AUTO CALIB. C.
,
BOTH
time and temperature.
AUTO.CALIB.C – automatic adjustment time interval (for scale with internal
adjustment)
Specyfying time interval, determining how often automatic adjustment process shall
be initiated.
13.3
Manual adjustment
13.3.1 Internal adjustment
Press
Cal
button
or
Содержание HRP 150