The easiest way is to slip the batteries one after another from the flat spring
side and next push them into the spiral spring direction. Finally the user
should screw the nut tight.
New batteries life is about 7 month long.
If the device is not in use for a longer period of time the batteries should be
taken out of the device. If not it may happen that they will spill and damage
the device.
Taking the batteries out does not result in the calibration data loss.
6. NOTES
The thermometer is a water-proof device, it has been tested before being put
up for sale. Its leakproofness is assured due to three sealing rings. Their
placement is presented in the picture beneath.
Two of the rings are to be found on each side of the casing, the third one is
placed inside the nut holding the head with the electrodes. While replacing the
batteries it is recommended to grease the ring with the silicone in order not to
be constrained to screw the nut with too great force. Before the screwing it
needs to be checked whether the ring is in the correct place, next the nut
must be screwed. If it is left loose, the inside of the device may leek and spoil
the device. Damage of this kind is not covered by the warranty. The
thermometer head cannot be heated with a temperature higher than 70
°
C due
to the material its casing has been made of. The casing is sensitive to
dissolvent.
7. ERRORS DESCRIPTION
HLP
Factory calibration loss, contact RADWAG company.
Lo
Low battery charge state, showing in turn with the
measurement result.
Batteries must be replaced (description is to be found
above).