Titra
Lab
®
TitraLab 870 Reference Manual
Page 71
Cable
resistance
A cable has a given length, therefore a given resistance. When the
measured resistance is low (below 50
Ω
), the cable resistance is not
negligible and must be taken into account.
Enter the cable resistance when:
•
measuring low resistances (below 50
Ω
) or high conductances
(above 20 mS).
•
using a 2 or 3-pole conductivity cell.
The cable resistance is normally specified by the manufacturer.
Cable resistances of a few Radiometer Analytical conductivity cells
are given below:
Figure 14: Cable resistances of Radiometer Analytical conductivity cells
If you create a conductivity cell from the Catalogue list, the
cable resistance is automatically assigned to the
conductivity cell created (and cannot be modified).
Enter 0 for the cable resistance of a 4-pole conductivity cell
(whatever the conductivity cell used).
Enter in:
When creating an electrode with the Conductivity function and the
option
From = Other
.
Refer to "Create electrode", page 87
Available limits:
0.000 to 9.999
Ω
Conductiivity cell
Cable resistance (
Ω
)
CDC511T
0
CDC861T
0
CDC565
0
CDC749
0.180
CDC267-9 with cable A94L136
0.145
CDC267-9 with cable A94L336
0.350
CDC241-9 with cable A94L136
0.145
CDC241-9 with cable A94L336
0.350
XE100 with cable A94L136
0.145
XE100 with cable A94L336
0.350
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...