The dBRL type option selects the method for dBRL
calculation. “Total” displays the dBRL of the fault, plus
the attenuation of the cable. “Fault” displays dBRL of
the fault
minus
the attenuation of the cable.
The horizontal reference option allows the user to
display a horizontal reference line on the center of the
display.
The backlight at start-up option is used to select
whether the LCD backlight is on or off at start-up.
The velocity format option selects whether the velocity
of propagation control is displayed as a percentage
of the speed of light (VOP) or as meters or feet per
microsecond velocity divided by 2 (V/2). VOP precision
selects the precision of the VOP display and can be
either 2 or 3 digit precision.
The cancel test lead length option allows the user to
automatically subtract the length associated with the
test leads from the distance between cursor readout.
The instrument will place the first cursor at the end of
the test leads.
NOTE: If test leads are not used, make sure to
disable the cancel test lead length.
The serial printer option is used to select the type of
serial printer for RS232 printing. The options available
are Seiko DPU 411 thermal printer and the Citizen PN60
plain paper printer.
The Auto Filter option enables or disables automatic
software noise filtering. When enabled, the software
noise filtering will automatically be activated if a high
level of external noise is detected on the input.
2.3.9 Waveform Storage and Recall
Model 1205CXA’s SUPER-STORE waveform storage
capability allows the operator to store a waveform for
later comparison and analysis. SUPER-STORE stores
the entire cable under test, not just the section of cable
displayed on screen at the time of storage. This feature is
helpful if: the incorrect section of cable was on screen at
the time of storage; comparing two separate waveforms
(cables); or for comparing the same waveform (cable)
before and after repairing the cable.
The Model 1205CXA comes standard with eight SUPER-
STORE waveform memory locations. The Extended
memory option is available to increase the stored
waveform capacity to 32 waveforms. The waveform(s)
will remain in storage, whether the instrument is on or
off.