Safety/ Operating Instructions - Technical Data

RLWE 4161

4. Scope of delivery

1x Basic unit 4-channel fiber-optic pulse receiver RLWE 4161

1x Mains cable 230 VAC, 2 m

1x Operating manual

5. Technical characteristics

Parameter

Min.

Typ.

Max.

Unit

Notes

Optical inputs

Input signal frequency (rectangular signal)

>100

MHz

Budget of optical input power

8

15

dB

Type of input receptacles:

FSMA (female)

TTL outputs

Output impedance

50

Ω

Output signal frequency

>100

MHz

Output level

TTL output voltage Low

0

0.7

1

V

TTL output voltage High

1.8

2.8

4.8

V

2)

Output signal rising and falling slopes
(L-H and H-L)

<1

ns

Propagation delay between optical inputs and 
TTL outputs

<55

ns

Skewing between two outputs of a channel

<1

ns

RMS additive-jitter at 13 MHz (1 Hz – 5 MHz)

<5

ps

Phase noise of outputs

5)

5)

Type of output receptacles:

BNC (female)

NIM outputs 

6)

Output impedance

50

Ω

Output signal frequency

>100

MHz

Output level

NIM output voltage Low at 50 Ω load

0.05

V

NIM output voltage High at 50 Ω load

-0.7

V

Output signal rising and falling slopes

1.5

ns

4)

RADIOCON, owner W. Kern, Bamberger Str. 7, D-01187 Dresden, Germany

T +49 351 32 33 549, E 

[email protected]

, www.radiocon.de

page

Data sheet 31S0009A

6 (9)

Содержание RLWE 4161

Страница 1: ...signal sources transmitting at nominal 850 nm wave length At rear side of this product input signals can be coupled into the device via FSMA fiber cable receptacles There are four 50 Ω output signal domains CH1 and CH3 with four TTL signal outputs each and two further output domains CH2 and CH4 for four NIM output signals which can be used in parallel Optical channels 1 and 2 and channels 3 and 4 ...

Страница 2: ... which are connected to this product This device shall treated carefully and shall not be dropped down because this product could be damaged by impacts or by falls from low levels This product shall neither exposed to extreme temperatures nor to humidity This device is designed for indoor use only especially in laboratories and has no spray water protection In case there is no safe operation possi...

Страница 3: ...1 Figure 2 Operating an indication elements on front panel of RLWE 4161 Figure 3 Interfaces on device rear side of RLWE 4161 with four fiber optic inputs RADIOCON owner W Kern Bamberger Str 7 D 01187 Dresden Germany T 49 351 32 33 549 E radiocon radiocon de www radiocon de page Data sheet 31S0009A 3 9 3 4 2 1 7 8 10 14 13 9 5 6 12 11 15 16 ...

Страница 4: ...nnel i 1 to electrical output channel i resp 8 NIM output receptacles 9 TTL signal indicator 10 Inverting TTL output with marking INV 11 Toggle switch for straight forward and crossed over signal mapping 12 NIM signal indicator 13 Inverting NIM output with marking INV 14 Operation indicator 15 Mains socket with mains power switch mains fuses and mains EMC filter 16 Four FSMA receptacles of fiber o...

Страница 5: ...hannels 3 and 4 in conjunction with electrical output channels CH3 and CH4 This way optical signal pairs 1 2 and 3 4 resp can be converted optionally into TTL or NIM signals Activated mapping modes are indicated by red LEDs above 6 and 7 resp on front panel As soon as a signal is available on a channel output the yellow signal LEDs 9 and 12 resp are glowing In case there is no signal available the...

Страница 6: ... 0 7 1 V TTL output voltage High 1 8 2 8 4 8 V 2 Output signal rising and falling slopes L H and H L 1 ns Propagation delay between optical inputs and TTL outputs 55 ns Skewing between two outputs of a channel 1 ns RMS additive jitter at 13 MHz 1 Hz 5 MHz 5 ps Phase noise of outputs 5 5 Type of output receptacles BNC female NIM outputs 6 Output impedance 50 Ω Output signal frequency 100 MHz Output...

Страница 7: ...equency 47 63 Hz Power consumption 36 W Fuses 2 x miniature fuse 5x20 mm 1 A 250 V slow Mechanical dimensions Width 483 mm Heigth 44 mm Depth 245 mm Total weight 2 kg Accessories 1 x Mains cable 2 m Notes 1 NA 0 27 max input power 24 dBm 2 Typical at loed of 50 Ω Maximum without load 3 At 50 Ω 4 Time difference between 10 and 90 of signal amplitude test signal frequency 10 MHz 5 Refer also diagram...

Страница 8: ...Hz test signals at RT RADIOCON owner W Kern Bamberger Str 7 D 01187 Dresden Germany T 49 351 32 33 549 E radiocon radiocon de www radiocon de page Data sheet 31S0009A 8 9 Figure 4 Typical phase noise of RLWE 4161 on TTL channels Figure 5 Typical phase noise of RLWE 4161 on NIM channels ...

Страница 9: ...ical scrap This way the material of this product can by recycled professionally and can be used best in a closed material cycle 7 Order instructions Product name RLWE 4161 Order address RADIOCON W Kern Bamberger Straße 7 D 01187 Dresden Germany email radiocon radiocon de Pulse and clock dividers also with customized device shapes with different digital input and output definitions or as electrical...

Отзывы: