Symbol Descriptions 

 

The CE mark and Notified Body Registration 
Numbers, the requirement of Annex II from 
Medical Device Directive 93/42/EEC are 
met. 

 

Please do not dispose of the product in 
the household waste at the end of its 
useful life. Disposal can take place at 
appropriate collection points provided in 
your country.

 

 

Do not reuse 

 

Caution 

 

Please read the instructions for use 

 

Paper Recycling 

 

Manufacturer 

 

To protect the environment, dispose of 
empty batteries at your retail store or at 
appropriate collection sites according to 
national or local regulations.

 

 

BF type applied 
part 

Authorized representative in the European 
community 

IP22 

Classification for water ingress and particulate matter. 

 
 

Manufacturer’s declaration-electromagnetic emissions

 

The TH7xy series is intended for use in the electromagnetic environment (for home healthcare) specified below. 

The customer or the user of the TH7xy series should assure that it is used in such an environment. 

Emission test 

Compliance Electromagnetic environment-guidance

 

(for home healthcare environment) 

RF emissions CISPR 11 

Group 1 

The TH7xy series uses RF energy only for its internal function. Therefore, its RF emissions 

are very low and are not likely to cause any interference in nearby electronic equipment. 

RF emissions CISPR 11 

Class B 

The TH7xy series is suitable for use in all establishments, including domestic establishments 

and those directly connected to the public low-voltage power supply network that supplies 

buildings used for domestic purposes. 

 

Manufacturer’s declaration-electromagnetic immunity 

The TH7xy series is intended for use in the electromagnetic environment (for home healthcare) specified below. 

The customer or the user of the TH7xy series should assure that it is used in such an environment.

Immunity test 

IEC 60601   

test level 

Compliance level 

Electromagnetic environment-guidance    

(for home healthcare environment) 

Electrostatic 
discharge(ESD) IEC 
61000-4-2 

Contact:±8 kV   
Air±2 kV,±4 kV,±8 kV,±15 kV 

Contact:±8 kV   

Air±2 kV,±4 kV,±8 kV,±15 kV 

Floors should be wood, concrete or ceramic tile. 
If floors are covered with synthetic material, the 
relative humidity should be at least 30% 

Power 
frequency(50, 60Hz) 
magnetic field IEC 
61000-4-8 

30 A/m 
50 Hz or 60 Hz 

30 A/m 

50 Hz and 60 Hz 

The TH7xy series power frequency magnetic 
fields should be at levels characteristic of a 
typical location in a typical home healthcare 
environment. 

 

Manufacturer’s declaration-electromagnetic immunity 

The TH7xy series is intended for use in the electromagnetic environment (for home healthcare) specified below. 

The customer or the user of the TH7xy series should assure that is used in such and environment. 

Immunity 

test 

IEC 60601 test 

level 

Compliance 

level 

Electromagnetic environment-guidance   

(for home healthcare environment) 

Radiated RF   
IEC 
61000-4-3 

10 V/m 
80 MHz – 2,7 GHz   
80 % AM at 1 kHz 

10 V/m 
80 MHz – 2,7 
GHz  
80 % AM at 1 kHz   

Recommended separation distance: 

d = 1,2 

√P

 

d = 1,2 

√P

   

80MHz to 800 MHz 

d = 2,3 

√P

    800MHz to 2,7 GHz   

Where

 P

 is the maximum output power rating of the transmitter in watts (W) 

according to the transmitter manufacturer and 

is the recommended 

separation distance in metres (m). 

Field strengths from fixed RF transmitters, as determined by an 

electromagnetic site survey, 

a

 should be less than the compliance level in each 

frequency range.

Interference may occur in the vicinity of equipment marked with the following 

symbol: 

 

NOTE1   At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE2      These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects and people. 
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, 

AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment 

due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the 

TH7xy series is used exceeds the applicable RF compliance level above, the TH7xy series should be observed to verify normal operation. If 

abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the TH7xy series. 

 

Recommended separation distance between 

portable and mobile RF communications equipment and the 

TH7xy series

 

The TH7xy series is intended for use in an electromagnetic environment (for home healthcare) in which radiated RF disturbances are 

controlled. The customer or the user of the TH7xy series can help prevent electromagnetic interference by maintaining a minimum distance 

between portable and mobile RF communications equipment (transmitters) and the TH7xy series as recommended below, according to the 

maximum output power of the communications equipment. 

Rated maximum 

output power of 

transmitter 

Separation distance according to frequency of transmitter 

150 kHz to 80 MHz 

d =1,2

√P

 

80 MHz to 800 MHz 

d =1,2

√P

 

800 MHz to 2,7 GHz 

d =2,3

√P

 

0,01 

N/A 

0,12 0,23 

0,1 

N/A 

0,38 0,73 

N/A 

1,2 2,3 

10 

N/A 

3,8 7,3 

100 

N/A 

12 23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

 in meters (m) can be estimated 

using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) 

according to the transmitter manufacturer. 
NOTE1    At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects and people.

 

Manufacturer’s declaration-electromagnetic immunity 

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment 

The TH7xy series is intended for use in the electromagnetic environment (for home healthcare) specified below. 

The customer or the user of the TH7xy series should assure that it is used in such an environment.

 

Test 

frequency 

(MHz) 

Band 

a)

 

(MHz) 

Service 

a)

 Modulation 

b)

 

Maximum 

power 

(W) 

Distance 

(m) 

IMMUNITY 

TEST LEVEL 

(V/m) 

Compliance LEVEL 

(V/m) 

(for home healthcare) 

385 380 

–390 

TETRA 400 

Pulse 

modulation b) 

18 Hz 

1,8 0,3  27 

27 

450 430 

– 

470 

GMRS 460, 

FRS 460 

FM c) 

±5 kHz deviation

 

1 kHz sine 

2 0,3  28 

28 

710 

704 – 787 

LTE Band 13, 

17 

Pulse 

modulation b) 

217 Hz 

0,2 0,3  9 

745 
780 

810 

800 – 960 

GSM 800/900, 

TETRA 800, 

iDEN 820, 

CDMA 850, 

LTE Band 5 

Pulse 

modulation b) 

18 Hz 

2 0,3  28 

28 

870 

930 

1 720 

1 700 – 

1 990 

GSM 1800; 

CDMA 1900; 

GSM 1900;   

DECT; 

LTE Band 1, 3, 

4, 25; UMTS 

Pulse 

modulation b) 

217 Hz 

2 0,3  28 

28 

1 845 

1 970 

2 450 

2 400 – 

2 570 

Bluetooth,  

WLAN, 

802.11 b/g/n,   

RFID 2450,   

LTE Band 7 

Pulse 

modulation b) 

217 Hz 

2 0,3  28 

28 

5 240 

5 100 – 

5 800 

WLAN 802.11 

a/n 

Pulse 

modulation b) 

217 Hz 

0,2 0,3  9 

5 500 
5 785 

NOTE      If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME 
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3. 
a)          For some services, only the uplink frequencies are included. 
b)          The carrier shall be modulated using a 50 % duty cycle square wave signal. 

c)          As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not  represent 

actual modulation, it would be worst case. 

 