Symbol Descriptions 

 

The CE mark and Notified Body 
Registration Numbers, the 
requirement of Annex II from Medical 
Device Directive 93/42/EEC are met.

 

 

Indicates this device is subject to the Waste 
Electrical and Electronic 
Equipment Directive in the European Union. 
To protect the environment, dispose of 
useless device at appropriate collection sites 
according to national or local regulations. 

 

Authorized 
representative in the 
European community 

 

Caution 

 

Please read the instructions for use 

 

Manufacturer 

 

Battery Recycling 

 

Paper Recycling 

IP22

 

Classification for water 
ingress and particulate 
matter. 

 

Guidance and manufacturer’s declaration – electromagnetic emissions

 

The TMS-51A is intended for use in the electromagnetic environment specified below. The customer or the user of the TMS-51A should assure that it is 
used in such an environment.

 

Emissions test

 

Compliance

 

Electromagnetic environment 

– 

guidance

 

RF emissions 
CISPR 11

 

Group 1

 

The TMS-51A uses RF energy only for its internal function. Therefore, its RF emissions are 
very low and are not likely to cause any interference in nearby electronic equipment. 

 

RF emissions 
CISPR 11

 

Class B

 

The TMS-51A is suitable for use in all establishments, including domestic establishments and 
those directly connected to the public low-voltage power supply network that supplies buildings 
used for domestic purposes.

 

Harmonic emissions 
IEC 61000-3-2

 

Not applicable

 

Voltage fluctuations/ 
flicker emissions 
IEC 61000-3-3

 

Not applicable

 

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The TMS-51A is intended for use in the electromagnetic environment specified below. The customer or the user of the TMS-51A should assure that it is 
used in such an environment. 

Immunity test 

IEC 60601 test level

 

Compliance 

level

 

Electromagnetic environment 

– 

guidance

 

Conducted RF 
IEC 61000-4-6 
 
 
Radiated RF 
IEC 61000-4-3 

3 Vrms 
150 kHz to 80 MHz 
 

 

3 V/m 
80 MHz to 2,5 GHz 

Not applicable 

 
 
 

3  V/m 

Portable and mobile RF communications equipment should be used no closer to any part 
of the TMS-51A, including cables, than the recommended separation distance calculated 
from the equation applicable to the frequency of the transmitter. 

 
Recommended separation distance

                                                                                                                                                                                           

d = 

1,2 

                                                                                                                                                                           

d = 

1,2 

  80 MHz to 800 MHz

                                                                                                                                                                                   

d = 

2,3 

  800 MHz to 2,5 GHz 

 
where 

is the maximum output power rating of the transmitter in watts (W) according to 

the transmitter manufacturer and 

is the recommended separation distance in meters 

(m). 
 
Field strengths from fixed RF transmitters, as determined by an electromagnetic site 
survey, a should be less than the compliance level in each frequency range b. 
 
Interference may occur in the vicinity of equipment marked with the following symbol: 

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM 

radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the TMS-51A is used exceeds 
the applicable RF compliance level above, the TMS-51A should be observed to verify normal operation. If abnormal performance is observed, additional 
measures may be necessary, such as reorienting or relocating the TMS-51A. 

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than  3  V/m. 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The TMS-51A is intended for use in the electromagnetic environment specified below. The customer or the user of the TMS-51A should assure that it is 
used in such an environment. 

Immunity test

 

IEC 60601 test level

 

Compliance level

 

Electromagnetic environment 

– 

guidance

 