background image

Ref No.

032012 

Ref No.

082012 

4.2 Measuring temperature under Surface mode:   

4.2.1 When you press the “SCAN” button, you will get the real time temperature immediately. If you press and hold the “SCAN” button, the reading of measurement 

will be continuous updated.   

4.2.2 Applications include temperature measurements for Water, Milk, Cloth, Skin or other object. 

* Note: This mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. 

4.3 Get the room temperature 

4.3.1 For room temperature, the thermometer should be placed on the table and avoid the direct sunshine or air conditioner flow on.   
4.3.2 When in power off, room temperature will always be shown on the screen with 

  icon. 

4.3.3 The room temperature can be reference after 15min later. 

5. After Measurement: 

5.1 Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life. The LCD will only show the room temperature. 
5.2 Clean the probe after each use to ensure an accurate reading and avoid cross contamination.    (See the section of Care and Cleaning for details.) 
5.3 Remember to put on the Probe Cap when not in use.   

 

Important Notes 

Care and cleaning

 

  T

he probe is the most delicate part of the thermometer. Use with care when cleaning the lens to avoid damage.   

a. After the measurement, please use the cotton swab with the Alcohol (70% concentration) to clean the lens( on the inside of the probe). 
b. Allow the probe to fully dry for at least 1 minute. 
c.

 

Storage temperature range: It should be stored at room temperature between -20~+50°C , RH

85%   

d. Keep the unit dry and away from any liquids and direct sunlight.   
e. The Probe should not be submerged into liquids. 

* Note: Please check the device if damaged once it falls. If you can’t make sure of it ,please send the complete device to       

the nearest retailer for recalibration.   

  Holding the thermometer too long may cause a higher ambient temperature reading of the probe.   

This could make the body temperature measurement lower than usual. 

Battery replacement 
 

 

1

  Open the battery cover: Insert a pointed object, such as a pen, into battery cover pick hole.   

At the same time, use thumb to push battery cover out.   

2

  Hold the device and flip the battery out with a small screwdriver   

3

  Insert  the  new  battery  under  the  metal  hook  on  the  left  side  and  press  the  right  side  of  the 

battery down until you hear a click   

4

  Replace the battery cover 

 

Keep the battery away from children.(This device is supplied with one lithium cell       

          (CR2032x1).)

 

 

 

The positive (+) side up and the negative (-) side pointed down.

   

 

 

Trouble shooting: 

Error Message 

Problem 

Solution 

 

Error 5~9, the system is not functioning properly. 

Unload the battery, wait for 1 minute and repower it. If the message 
reappears, contact the retailer for service. 

 

Measurement before device stabilization.

 

Wait until all the icons stops flashing.

 

 

The device showing a rapid ambient temperature change.

 

 

The ambient temperature is not within the range between 10

o

C~40

o

C (50

o

F ~104

o

F). 

Allow the thermometer to rest in a room for at least 30 minutes at room 
temperature: 10

o

C and 40

o

C (50

o

F ~104

o

F).

 

 

(1)

 

In Ear mode: Temperature taken is higher than +42.2

o

C (108°F) 

(2)

 

In Surface mode: Temperature taken is higher than +80

o

C (176°F) 

 

(1)

 

In Ear mode: Temperature taken is lower than +34

o

C (93.2°F) 

(2)

 

In Surface mode: Temperature taken is lower than –22

o

C (-7.6°F) 

Please select the target within specifications. If a malfunction still exists, 
please contact the nearest retailer. 

 

Device cannot be powered on to the ready stage. 

Change with a new battery. 

 

Warranty:

 

Warranty :12 months   
Manufacture Date

as the serial number( please open the battery cover, it is shown on the inside of the device.)

 

 

Ex.SN:E912A000001, the first “E” is External, the second number “9” is the manufacture year 2009, the third and the fourth number “12” is the manufacture month, the 
others is the serial number. 

Note: The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy of temperature measurements, please contact the 

dealers or nearest service address. 

 

Please read the instructions for use       

  BF type applied part 

 

Radiant Innovation Inc.    Http://www.radiantek.com.tw 
Add: 1F, No.3, Industrial East 9

th

 Road, Science-Based Industrial Park, 

HsinChu, Taiwan 300. 

Medical Technology Promedt Consulting GmbH   
Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany 

Отзывы: