TH09F Non-contact Forehead Thermometer

 

 

Functions 

Forehead temperature 

The thermometer has been designed for  home use. It’s not meant to replace a visit to the doctor. 
Please also remember to compare the measurement result to your regular body temperature.

 

Please 

consult with doctor if you have health concerns. 

 Please see the “Use of the thermometer” section to learn how to measure the body temperature. 

 

Fever indication 

If the thermometer detects a temperature ≥37.5°C (99.5°F) under forehead mode, three short beep 
sounds will follow one long beep sound to warn the user for potential fever. 
If the thermometer detects a temperature ≥ 38°C (100.4°F), the “Fever Indicator” will be "Red" light 
signal. 

RED: Continually monitor the body temperature. If in doubt about the person’s condition, seek   

medical assistance. 

 

Memory locations 

There is 1 set memory for body temperature. 

 After  measuring the body temperature, the  device is turned on again and the last reading with 

  icon is displayed on the LCD. 

 

°C / °F switch 

In “

Power Off

” mode, press and hold the “START” button.

 

After about 8 seconds, the display will 

show “°C” flashing. Release "START" button and press again and release. The display will show "°F" 
flashing. After about 5 seconds, the display will auto reset and there will be 2 short beeps. You can 
also use the same process to change the LCD display from °F to °C.

 

 

 

Use of the thermometer

 

Note: If there is any temperature difference between the places where the device is stored and where you are going to measure, subject and the device should   

stay in the same room for at least 15 minutes before measurement. 

1. Always make sure the probe lens is clean without any damage and the forehead is clean. 

 

Warning: Choking from swallowing small parts and batteries by children or pets is possible, please keep small parts and batteries at places where   

children and pets can’t reach. 

2. Power on: 
   

Press the “START” button.

   

 

3. Measuring body temperature on the forehead:

 

    Press the “START” button to power on the device. You can see the 

  icon on the screen and hear two beep sounds. In this mode, you can hold 

the thermometer within 3 cm from the central forehead (Fig. 1) and press the "START" button to get the forehead measurement. The time 
consuming for measurement might be 1 second. After two short beeps are heard, the 

  icon stops flashing to be ready for next measurement. 

         

Figure 1

 

 

Points for attention:

   

a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its “oral-equivalent” value. 

b. Before the measurement, the subject should stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins. 

                c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature. 

d.

 

The “Clinical Bias” is -1.4 ~ -1.7°C. 

e. The “Limits of Agreement” is 0.98. 

f. The "Repeatability" is 0.20°C. 

 

4. Power off: 

Device will automatically shut off if left idle for more than 1 minute to extend battery life. 

Intended use: 

The non-contact forehead thermometer is electronic thermometer using an 
infrared detector (thermopile detector) to detect body temperature from forehead 
center in people of all ages.

 

 
 

 

Intended operator:

 

The  intended  operator  should  have  at  least  eight  years  of  education,  including 
patients or other operators.. 

Specifications: 

 

Temperature measurement range: 28~43°C (82.4~109.4°F) 

 

Operating temperature range: 10~40°C (50~104°F), 15%~85% RH 

 

Storage temperature range: It should be stored at room temperature between 

-20~+50°C, RH≤85% 

Transportation temperature shall be less than 70°C, RH≤95% 

 

Atmospheric pressure: 800~1013 hPa 

 

Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC), 

IEC/EN60601-1(Safety) standards, ISO10993, RoHS. 

 

Accuracy: 

±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range. 

 

Fever indication, memory and °C/°F switch function 

 

Battery: one lithium cell battery (CR2032 *1pcs).   

 

Battery life: around 3,000

 

continuous readings. 

 

Expected Service Life: 4 years 

 

This thermometer is an adjusted mode thermometer that converts the forehead 

temperature to display its “oral equivalent.”   
(according to the result of the clinical evaluation to get the offset value) 

 

Enclosure Rating: IP22 

 

Dimensions: 102.98 x 34.98 x 22.70 mm 

 

Weight: 48.4 grams including battery 

 

The device should not submerge into any liquids and expose it to direct moisture.

 

 

There is no gender and age limitation for using the infrared thermometer. 

  This is not an AP

 

or APG product.

 

Содержание TH09F

Страница 1: ...m from the central forehead Fig 1 and press the START button to get the forehead measurement The time consuming for measurement might be 1 second After two short beeps are heard the icon stops flashing to be ready for next measurement Figure 1 Points for attention a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent value b Bef...

Страница 2: ...it for Er1 to disappear The ambient temperature is not within the range between 10o C and 40o C Allow the thermometer to rest in a room for at least 15 minutes at room temperature 10o C and 40o C Temperature taken is higher than 43o C Please select the target within specifications If a malfunction still exists please contact the nearest retailer Temperature taken is lower than 28o C Device cannot ...

Страница 3: ...ic tile If floors are covered with synthetic material the relative humidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 30 A m 50 Hz or 60 Hz 30 A m 50 Hz and 60 Hz The TH09F power frequency magnetic fields should be at levels characteristic of a typical location in a typical home healthcare environment Manufacturer s declaration electromagnetic immunity The TH09F i...

Страница 4: ...er s declaration electromagnetic immunity Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment The TH09F is intended for use in the electromagnetic environment for home healthcare specified below The customer or the user of the TH09F should assure that it is used in such an environment Test frequency MHz Band a MHz Service a Modulation b Maximum power W Distance ...