background image

 

Recommended separation distances between 

portable and mobile RF communications equipment and the 

TH0xyz series

 

The 

TH0xyz series

  is intended for use in an electromagnetic environment (for home healthcare) in which 

radiated RF disturbances are controlled. The customer or the user of the 

TH0xyz series

  can help prevent 

electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the 

TH0xyz series

  as recommended below, according to 

the maximum output power of the communications equipment.

 

Rated maximum 

output power of 

transmitter 

Separation distance according to frequency of transmitter 

m

 

150 kHz to 80 MHz 

d = 

1,2 

 

80 MHz to 800 MHz 

d = 

1,2 

 

800 MHz to 2,7 GHz 

d = 

2,3 

 

0,01

 

N/A 0,12 0,23 

0,1

 

N/A

 

0,38

 

0,73

 

1

 

N/A

 

1,2

 

2,3

 

10

 

N/A

 

3,8

 

7,3

 

100

 

N/A

 

12

 

23

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where 

is the maximum output power rating of the transmitter in watts (W) according to the transmitter 
manufacturer.  
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

 

 

Manufacturer’s declaration-electromagnetic immunity 

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment

The 

TH0xyz series

  is intended for use in the electromagnetic environment (for home healthcare) specified 

below.The customer or the user of the 

TH0xyz series

  should assure that it is used in such an environment.

Test 

frequency 

(MHz) 

Band 

a)

 

(MHz) 

Service 

a)

Modulation 

b)

 

Maximum 

power 

(W) 

Distance

(m) 

IMMUNITY 

TEST LEVEL

(V/m) 

Compliance 

LEVEL 

(V/m) 

(for home 

healthcare)

385 

380

390  TETRA 400

Pulse 

modulation b) 

18 Hz 

1,8 0,3  27 

27 

450 

430

470 

GMRS 460,

FRS 460

FM c) 

±

5 kHz 

deviation 

1 kHz sine 

2 0,3  28  28 

710 

704

787

LTE Band 

13,17 

Pulse 

modulation b)

217 Hz 

0,2 0,3  9 

745 

780 

810 

800

960

GSM 

800/900,

TETRA 

800, 

iDEN 820,

CDMA 850,

LTE Band 5

Pulse 

modulation b)

18 Hz 

2 0,3  28  28 

870 

 

930 

1720 

1700

 

1990 

GSM 1800;

CDMA 

1900;

GSM 

1900; 

DECT;

LTE Band 

1, 3, 

4, 25; 

UMTS

Pulse 

modulation b)

217 Hz 

2 0,3  28  28 

1845 

1970 

2450 

2400

 

2570 

Bluetooth, 

WLAN, 

802.11 

b/g/n, 

RFID 2450, 

LTE Band 7

Pulse 

modulation b)

217 Hz 

2 0,3  28  28 

5240 

5100

 

5800 

WLAN 

802.11 

a/n 

Pulse 

modulation b)

217 Hz 

0,2 0,3  9 

5500 

5785 

NOTE    If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting

antenna  and the ME EQUIPMENT or ME  SYSTEM may be reduced to 1 m. The 1 m test 
distance is permitted by IEC 61000-4-3. 

a)          For some services, only the uplink frequencies are included. 
b)          The carrier shall be modulated using a 50 % duty cycle square wave signal. 

c)      As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because

while it does not  represent actual modulation, it would be worst case. 

 

 

Отзывы: