Do not use battery not provided by RADIANQBIO
RADIANQBIO is not responsible for damage/malfunction in
operation caused by unofficial battery use. Uses of unofficial
battery may cause serious damage / malfunction on AED
Attach the AED pads clearly on the patient’s bare chest
For accurate ECG analysis and proper electric shock
delivery, AED pads shall be attached on the patient’s bare
chest firmly and tightly. Please refer to the pads location
according to patient's population, adult or child.
There should
be no air bubble between pads and applied site. Users shall
make sure the patient’s skin is clear and dry prior to the AED
pad attachment. Do not place the pads close to each other.
During ECG analysis, stay away from the patient
Touching the patient during the ECG analysis phase can
cause interference with the diagnostic process that may cause
increased analysis time, or result analysis error. Avoid
contact with the patient while analysis is on progress
AED pads are single-use only
• Do not reuse AED pads. This pouch bag contains
disposable pads. Do not open the package before use. If the
sealed package is damaged or opened, exchange it to new
one immediately.
• When the package was opened, the pads should be used
immediately and replace the pads every 24 hours for long
term use.
• Do not fold or apply excessive force to the package.
Interference may be caused by implanted pacemaker
With some pacemakers, the AED may not advise an electric
shock
• Do not attach the AED pads directly on an implanted device
• Attach the AED pads at least 5cm away from any implanted
device
Temperature requirement for stable operation/storage
Main device of HR-501, battery and electro-pads
recommend to maintain between 0˚C ~ 50˚C.
In case of shipping/transporting(within 5days) condition will
be maintain between -20˚C ~ 60˚C(within 5days), If it is out
of 0˚C ~ 50˚.